Organic pollution monitor UV meter OPM-161L-OPM-161L
Organic pollution monitor UV meter OPM-161L-Oki Riko Jizai Co., Ltd.

Organic pollution monitor UV meter OPM-161L
Oki Riko Jizai Co., Ltd.

Oki Riko Jizai Co., Ltd.'s Response Status

Response Rate

100.0%

Response Time

14.5hours

Relatively Fast Response


About This Product

■Summary

This is a measuring instrument that uses ultraviolet absorption photometry to determine the degree of organic pollution in water from the absorbance of ultraviolet light. Environmental impact has been reduced by using a mercury-free LED light source. The detector is a small, lightweight, straight-shaped direct immersion type, and can also be installed by being dropped.

■Measurement range

UV absorbance: 0~2.5Abs VIS absorbance: 0~1.0Abs Measured values ​​can also be displayed as absorbance converted to 10mm cell length.

■UV absorbance (when converted to 10mm cell ON)

・25mm cell: 0~1.0Abs ・10mm cell: 0~2.5Abs ・6mm cell: 0~4.17Abs

■VIS absorbance (when 10mm cell conversion is ON)

・25mm cell: 0~0.4Abs ・10mm cell: 0~1.0Abs ・6mm cell: 0~1.67Abs Cell cleaning method

■Automatic cleaning with wiper

・Cleaning cycle: 0 to 9,999 minutes (default setting is 10 ・Number of washes: 0 to 99 times (default setting is 1) ・Waiting time before cleaning: 0 to 999 seconds (default setting is 0) ・Waiting time after cleaning: 0 to 999 seconds (default setting is 0)

■Transmission output

DC 4 to 20mA Load resistance 600Ω or less Insulated type Select any 3 items from UV-VIS absorbance / UV absorbance / VIS absorbance / COD conversion value / Turbidity conversion value / SS concentration conversion value / Sample water temperature Simultaneous output possible (non-isolated between transmission outputs) COD conversion value, turbidity conversion value, and SS concentration conversion value are values ​​obtained by converting the absorbance into a linear equation. Contact output signal

■Alarm signal a (NO) Contact output: 1 power failure output and any 3 points below can be selected

・Under maintenance ・Cleaning in progress ・Detector water leakage ・UVLED abnormality ・VISLED abnormality ・Motor abnormality ・Abnormal sample water temperature ・Measurement value upper limit alarm ・Batch alert ・Power off signal c (NO/NC) Fixed to contact output 4 ・Contact capacity DC 30V 0.1A or less AC 125V 0.1A or less (resistive load)

■Contact input signal

・Cleaning start signal...by external program ・For cleaning control ・No-voltage contact input ・ON resistance 50Ω or less, short circuit current max. 5mA, ・Open voltage DC 24V, make time 0.1 seconds or more

  • Product

    Organic pollution monitor UV meter OPM-161L

Share this product


50+ people viewing

Last viewed: 2 hours ago


Free
Get started with our free quotation service - no cost, no obligation.

No Phone Required
We respect your privacy. You can receive quotes without sharing your phone number.

1 Models of Organic pollution monitor UV meter OPM-161L

Image Part Number Price (excluding tax) Power supply Related equipment Mass Display items Displayable items Measuring range Measurement method Measurement target Measurement cell Power consumption Detector maximum water depth Calibration method Dedicated cable External dimensions Converter wiring port Paint color Installation method Protective structure Main material Manufacturer
Organic pollution monitor UV meter OPM-161L-Part Number-OPM-161L

OPM-161L

Available upon quote

AC 100~240V±10% 50/60Hz

Load amount calculator: When calculating the pollution load amount, we recommend the load amount calculator model CALD-2030.

Converter: Approx. 2.0kg (not including mount weight)
Detector: Approx. 3.0kg (excluding cable mass)

UV-VIS absorbance/UV absorbance/VIS absorbance/COD conversion value/turbidity conversion value/SS concentration conversion value/sample water temperature

UV-VIS absorbance / UV absorbance / VIS absorbance / COD conversion value / Turbidity conversion value / SS concentration conversion value / Sample water temperature

UV absorbance (0~2.5Abs) VIS absorbance (0~1.0Abs)

Dual wavelength absorption photometry (UV…255nm/VIS…660nm)

Organic pollutants in water

Immersed parallel cell

Average approximately 10VA, maximum approximately 20VA

6m

Zero calibration: using pure water, etc.
Span calibration:
UV…by potassium hydrogen phthalate solution
Turbidity…depending on formazin solution
(The calibration container is a standard accessory.)

Detector ~ 1 converter Standard length; Water sampling type...3m
Immersed type, drop-in type, hanging type, hoisting type (use of existing hoisting device only)...5m or 10m
Maximum specified length 30m (optional)

Converter; approx. 181 (W) × 180 (H) × 95 (D) mm

6 cable glands (for outer diameter φ6~φ12 cable) (1 location is for converter ~ detector cable) Cable gland can be removed to connect conduit (G 1/2 x 6)

Converter: metallic silver

Immersion type, drop-in type, hanging type, water sampling type, hoisting type (using existing hoisting device)

Converter: IP65 Detector: IPX

Converter: Aluminum die-casting
Detector: Main body...SUS316, SUS316L, sapphire glass, FPM, zinc, silicone rubber
Cell window…sapphire glass
Wiper…FPM or SUS304

Toa DKK Co., Ltd.

Customers who viewed this product also viewed

Reviews shown here are reviews of companies.

See More Water Quality Measuring Instruments Products

Other products of Oki Riko Jizai Co., Ltd.

Reviews shown here are reviews of companies.


View more products of Oki Riko Jizai Co., Ltd.

About Company Handling This Product

Response Rate

100.0%


Response Time

14.5hrs

  • Japan

This is the version of our website addressed to speakers of English in the United States. If you are a resident of another country, please select the appropriate version of Metoree for your country in the drop-down menu.

Copyright © 2024 Metoree