High-speed, compact gas analyzer Micro GC Agilent 490-Agilent 490
High-speed, compact gas analyzer Micro GC Agilent 490-Oki Riko Jizai Co., Ltd.

High-speed, compact gas analyzer Micro GC Agilent 490
Oki Riko Jizai Co., Ltd.

Oki Riko Jizai Co., Ltd.'s Response Status

Response Rate

100.0%

Response Time

14.5hours

Relatively Fast Response


About This Product

■Summary

Equipped with a high-sensitivity µTCD, achieving a detection limit of 1ppm. Pressure can be set from a PC using an electronic column pressure control valve. Control/monitoring is possible from a remote location via LAN connection.

■Features

・Equipped with a high-sensitivity µTCD, achieving a detection limit of 1ppm. ・The device stabilizes in about 15 minutes after startup. ・Pressure can be set from a PC using an electronic column pressure control valve. ・Preparing for heat injection. (Optional) ・Reduce measurement time with 5 types of backflush modules.

  • Product

    High-speed, compact gas analyzer Micro GC Agilent 490

Share this product


40+ people viewing


Free
Get started with our free quotation service - no cost, no obligation.

No Phone Required
We respect your privacy. You can receive quotes without sharing your phone number.

1 Models of High-speed, compact gas analyzer Micro GC Agilent 490

Image Part Number Price (excluding tax) Power supply/power consumption Weight Means of communication Installation location Measurement sample Measuring concentration range Detection limit Detector Size Analysis time Reproducibility Manufacturer Carrier gas Gas sampler Oven
High-speed, compact gas analyzer Micro GC Agilent 490-Part Number-Agilent 490

Agilent 490

Available upon quote

AC90~130V 50-60Hz 12VDC, maximum 130W (Quad 180W)

5.2kg (When using 2 channels, it varies depending on the number of built-in channels.)

LAN (RJ45), RS232C

Temperature range 0~50℃
Humidity range 0~95% R.H. (non-condensing)

Do not condense.
Temperature +5~40℃
Pressure 0~100kPa

1ppm~100% (Detection limit varies depending on conditions)

1ppm n-Pentane (using WCOT column, R.T. 15 seconds)
10ppm ethane (using micro-packed column at R.T. 25 seconds)

Thermal conductivity detector (µTCD)

150 (W) × 300 (D) × 280 (H) mm
Quad type is 550 (D) mm

600 seconds

R.S.D 0.5% or less (Propane at 1mol% level, using WCOT column, constant temperature, constant pressure)

GL Sciences Co., Ltd.

He, N2, Ar (supply pressure 550±10kPa) purity 99.995% or more

Injection volume 1~10μL (0~999msec) Can be set arbitrarily

30~180℃ (constant temperature analysis)

Customers who viewed this product also viewed

Reviews shown here are reviews of companies.

See More Gas Analyzers Products

Other products of Oki Riko Jizai Co., Ltd.

Reviews shown here are reviews of companies.


View more products of Oki Riko Jizai Co., Ltd.

About Company Handling This Product

Response Rate

100.0%


Response Time

14.5hrs

  • Japan

This is the version of our website addressed to speakers of English in the United States. If you are a resident of another country, please select the appropriate version of Metoree for your country in the drop-down menu.

Copyright © 2024 Metoree