All Categories
History
Response Rate
100.0%
Response Time
14.2hours
Relatively Fast Response
Product
Instrumentation system equipment Oxygen concentration meter/temperature/hygrometer Low concentration zirconia tabletop oxygen concentration meter OX400Handling Company
Yonei Co., Ltd.Categories
Image | Part Number | Price (excluding tax) | Power supply | Installation environment | Display | Self-diagnosis function error (failure) | Self-diagnosis function alarm (warning) | Measuring range | Gas to be measured | Measurement principle | Measurement gas conditions | Power consumption | Proofreading | Minimum resolution | Warm-up time | Connection terminal | Maximum allowable contact current | Maximum allowable contact voltage | Contact maximum power | Contact output | Contact input | Rated control capacity | Safety | External output | External dimensions/mass | Coating | Unit | Output contact specifications | Transmission output | Gas piping | Gas pressure | R.C.M. |
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
OX400 |
Available upon quote |
Rating: 100-120VAC/200-240VAC, 50/60HZ |
For indoor installation only in non-explosion-proof areas |
Oxygen concentration display: 4 digital digits (automatic switching) |
Sensor abnormality, heater temperature abnormality, temperature sensor disconnection, device temperature abnormality, CPU abnormality, FAN stop |
Heater instability, sensor failure, electromotive force abnormality, asymmetric voltage abnormality, calibration abnormality, sensor resistance value abnormality, oxygen concentration upper and lower limits, range over |
0-10ppm O2 ~ 0-100vol% O2 |
Oxygen concentration in inert gas |
Concentration battery type zirconia |
Flow rate: 200 ± 25ml/min (sensor part only) |
Max. 200VA (100-120VAC), Max. 400VA (200-240VAC) |
・All calibration (3 points): 10ppm, 1,000ppm, Air |
0.01ppm O2 |
Within 20 minutes |
External output terminal: M3 screw |
2A DC/AC |
30V DC, 264V AC (132V AC for 100V equipment) |
60W, 480VA |
Error contact, alarm contact (oxygen concentration upper/lower limit alarm contact), range marker contact |
1 point: Remote ON/OFF of measurement gas suction pump |
Resistive load 2A 30V DC, 2A 240V AC (120V AC for 100V equipment) |
CE, UKCA: EN 61010-1 |
First output: 4-20mA DC (maximum load resistance 550Ω) |
213 (W) × 132 (H) × approximately 375 (D) mm |
Polyester paint |
%, ppm |
No-voltage contact |
Communication method: unidirectional |
Gas inlet: Rc1/4 or 1/4NPT female thread |
Pump and aspirator method: 0 ~ 300PaG |
EN61326-1 Class A Table 2 |
Reviews shown here are reviews of companies.
Reviews shown here are reviews of companies.
Response Rate
100.0%
Response Time
14.2hrs