History
Product
Mini fixed point cell temperature furnace 9260Handling Company
Fluke CalibrationCategories
Image | Part Number | Price (excluding tax) | Power specifications | Weight | Communication | Environmental operating range | Temperature range | Degree of immersion | Comparison block | Fault protection | Duration of stable phase | Stability | Time to stability | External dimensions (height x width x depth) | Vertical slope | Heating time | Resolution | Cooling time | Preheat well | Display accuracy | Display scale | Well-to-well gradient (within comparison block) |
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
![]() |
9260 |
Available upon quote |
115 VAC (±10%), 60 Hz, 11 A, or |
20.5 kg (45 lb.) including block |
Built-in RS-232 |
5℃~45℃ |
50℃ ~ 680℃ |
229 mm (9 in) |
Two blocks with multiple holes, blank blocks, and custom blocks available |
Sensor burnout and short circuit protection, overtemperature cutout |
6 to 10 hours (typical) |
±0.03℃ up to 300℃ |
Nominal 15 minutes |
489 x 222 x 260mm |
Upper and lower zones can be adjusted by offset |
1.25 hours (25℃~680℃) |
0.01° |
10.5 hours (680℃~100℃) |
2 |
±0.2℃~300℃ |
Switchable between °C and °F |
±0.02℃ |