All Categories
History
Product
High compliance multi-product calibrator 5080AHandling Company
Fluke CalibrationCategories
Image | Part Number | Price (excluding tax) | Altitude | Power supply | Weight (without options) | Relative humidity | Temperature performance | Temperature coefficient | Power consumption | Standard interface | Resistance | Size | Safety | Frequency | Phase | Specification confidence level | Setting time | Warm up time | Analog low side isolation | EMC | DC current | DC voltage | DC power | AC current | AC voltage | AC power |
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
![]() |
5080A |
Available upon quote |
Operating: up to 2,000m |
Power supply voltage (selectable): 100V, 120V, 220V, 240V |
22kg |
Operating: <80% ~ 30℃, <70% ~ 40℃, <40% ~ 50℃ |
Operating: 0℃ ~ 50℃ |
The temperature coefficient for temperatures outside the tcal ±5°C range is 10% of the stated specification per 1°C from 0°C to 35°C. At temperatures above 35°C, it is 20% of the stated specification per 1°C. |
600VA |
RS-232 and Ethernet |
Range: 0 ~ 190 MΩ, 19 resistance values (x1 and x1.9) |
Height: 19.3 cm |
Compliant with EN. 61010-1:2001, CAN/CSA-C22.2 No. 61010-1-04, UL 61010-1: 2004 |
Range: 45Hz ~ 1kHz |
Range: 0 ~ ± 179.99° |
99% |
5 seconds or less for all functions unless otherwise specified |
Twice the time from power off to power on again, up to 30 minutes |
20V |
EN613261:2006 compliant |
Range: 0 ~ ±20.5A |
Range: 0 ~ ±1020V |
Range: 10.9 mW ~ 20.5 kW |
Range: 29 mA ~ 20.5A |
Range: 1 mV ~ 1,020V |
Range: 10.9 mW ~ 20.5 kW |
Reviews shown here are reviews of companies.
Reviews shown here are reviews of companies.