All Categories
History
Response Rate
100.0%
Response Time
1.8hours
Very Fast Response
Product
Physics and Chemistry Equipment Oxygen Measurement System Entry Level Photosynthesis/Respiration Research Gas Phase Oxygen Electrode System Leafview1Handling Company
Asahiko Trading Co., Ltd.Categories
Usage Scenarios
Chemical Industry UseImage | Part Number | Price (excluding tax) | Power supply | Electrode type | Electrode output | Oxygen consumption | Suitability | Measuring range | Temperature control | Residual current | Material | Response time | Dimensions (HWD) | Size | Resolution | Polarization voltage | Optical port | Signal input | Back off | Sample chamber | Sample area | Gain | Analog output | Oxylab+ Electrode Control Unit Weight | LD1/2 electrode chamber weight |
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Leafview1 |
Available upon quote |
95~260 universal input mains power, output 12VDC 2.5A |
Clark type polarographic oxygen sensor |
Typical 1.6μA at 21% O2 |
Typically <0.015µmol/hr^-1 |
Gas phase respiration/photosynthesis |
Oxygen: 0~100% |
Water jacket that can be connected to a separate constant temperature water tank |
Typical 0.04µA in 0% O2 |
Black acetal |
10~90%, typically less than 5 seconds |
90×135×85mm |
95 (D) × 75 (H) mm |
10 x 10-6 μmols/ml at 20℃ |
Selectable between 0.4~0.9V (0.7V is recommended by default) |
Cast acrylic top window |
S1 O2 electrode (SMB) |
Back off signal in 1mV steps |
Leaf chamber (7.5cc) |
10cm2 |
Coarse adjustment ×1, ×2, ×5, ×10, ×20, ×50, ×100 |
0~5V O2 signal |
320 grams |
350 grams |
Reviews shown here are reviews of companies.
Reviews shown here are reviews of companies.
Response Rate
100.0%
Response Time
1.8hrs