Optical measurement equipment Light source Radiant flux standard light source RS-10DHandling Company
Asahiko Trading Co., Ltd.Image | Part Number | Price (excluding tax) | Power supply | Current accuracy (long term) | Current output | Current lamp on/off switching time | Brightness *RS-70-2 diffuser included | Setting accuracy | Color temperature *RS-70-2 diffuser included | Color temperature (CCT) | Illuminance *Distance: 25cm | Temperature drift (after 8 minutes warm-up) | Temperature drift | Calibration interval | Calibration wavelength range@wavelength interval | Uniformity of radiation surface *Distance: 25cm | Radiance *RS-70-2 diffuser included | Irradiance NIST standard uncertainty | Dimensions/weight | Operating humidity | Operating temperature | Output port uniformity | Regulator type | Regulation | Diffuser uniformity |
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
![]() |
RS-10D |
Available upon quote |
100~240VAC, 50/60Hz max. 30W |
≦0.5% |
Max 4A |
Approximately 30 seconds |
850cd/m2 |
≦0.2% |
2,575±50K |
2,856±20K |
592lx (=55fc) |
≦0.1% |
≦±0.25%/10℃ |
W/cm2*nm |
300~1,100nm@5nm |
±2% (within a range of 50mm or more) |
W/cm2*sr*nm |
±2.5% |
L197mm×W175mm×H203mm 2.8kg |
10~85% (non-condensing) |
15~35℃ |
±0.75% |
Constant current |
±0.2% or less for 10V line change |
±3% (within a range of 25cm or more) |