Product
CO2 analyzer 0-10% S158-PHandling Company
Senecom Co., Ltd.Categories
Product Image | Part Number | Price (excluding tax) | Accuracy | Analog output, low sensitivity | Calibration adjustment | Current requirements | Dimensions H×W×D | Gas sampling mode | Humidity range | Low sensitivity setting output (linear) | Maximum gas flow | Measuring range (LCD display) | Operating temperature range | Power requirements | Pressure dependence | Reproducibility | Response time | Storage temperature range | Temperature controller | Warm-up time (@22℃) | Weight | Working principle |
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
S158-P |
Available upon quote | 0.1% of full scale | 0~10% | Zero and span | Average 175mA, Peak 450mA | 17×40×30 (cm) | Flow gas flow, closed chamber | 5~95%RH, non-condensing (dry gas flow recommended) | 0~10% at 0~5VDC | 650mL/min | 0~10.00% | 0~50℃ | 12VDC via 120VAC/60Hz adapter | Less than ±1% of reading | Approximately 30 seconds to reach final value of 95% at 250mL/min | -40~70℃ | Maximum setting 50℃ | Approximately 5 minutes | 4.7kg | Non-dispersive infrared sensor |