Constant temperature (constant humidity) room walk-in chamber E seriesHandling Company
ESPEC CORP.Categories
Image | Part Number | Price (excluding tax) | Power supply | Operable ambient temperature and humidity | Temperature control (humidity control) method | Space temperature (humidity) temperature deviation | Air conditioner unit | Humidity range | Temperature range | Temperature change rate Falling time | Temperature change rate Rise time | Temperature (humidity) temperature fluctuation | Temperature (humidity) gradient | Standard equipment | Insulation material | Door | Floor load capacity | Exterior material | Refrigerant | Freezing method | Interior material |
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
E series TBE |
Available upon quote |
AC200V 3φ 3W 50/60Hz (within ±5% of rated voltage) |
5~40℃/up to 75%rh |
Balanced temperature and humidity control system (BTHC system) Water vapor partial pressure control system |
2.5℃/8%rh |
Blower, heater, humidifier, refrigerator, evaporator, temperature detection end, humidity detection end |
10~95%rh (at+10~+80℃) |
−40~+80℃ |
0.4℃/min |
1℃/min |
±0.5℃/±4%rh |
2.5℃/8%rh |
Door observation window (W180 x H289mm), cable hole φ50mm, interior light (LED), |
Rigid urethane foam |
Single door W850×H1,800mm |
Equally distributed load: 6kPa (600kgf/m2) |
Color steel plate |
R-404A [Can be changed to low GWP refrigerant R-449A] |
Single-stage refrigeration air-cooled condenser or water-cooled condenser |
Stainless steel plate |