Sulfur carbon analyzer ■Power capacity Single phase 200 V +10%, -15% (boosted to 230 V), 50/60 Hz, 5 AA (idling), 18 A (maximum), heat output: 3,100 BTU/hr (idling) , 5,100 BTU/hr (at analysis) *Average when operating under normal settings.
You can search for other models from each index. The displayed value is the value of the currently selected part number.
Analysis accuracy
Analysis range (sample weight 200 mg)
Analysis time
Combustion furnace
Control
Detection method
Gas used
Operating environment
Reagent
Sample weight
Size
Weight
Part Number
SC832AHandling Company
LECO Japan LLCImage | Price (excluding tax) | Analysis accuracy | Analysis range (sample weight 200 mg) | Analysis time | Combustion furnace | Control | Detection method | Gas used | Operating environment | Reagent | Sample weight | Size | Weight |
---|---|---|---|---|---|---|---|---|---|---|---|---|---|
![]() |
Available upon quote | Sulfur: Standard deviation (s) 0.004 mg or RSD 1% *Whichever is greater Carbon: Standard deviation (s) 0.0088 mg or RSD 1% *Whichever is larger | Sulfur: 0.008mg~30mg (0.004%~15%) Carbon: 0.0175mg~200mg (0.009%~100%) | 60~120 seconds | 800℃~1,550℃++, set value ±10℃ | PC Windows® based software for CHN828 | Non-dispersive infrared absorption (NDIR) | Combustion: Oxygen 99.5%, 0.28 MPa ±10% | Temperature: 15℃~35℃ Humidity: 20%~80% (non-condensing) | Magnesium perchlorate (Anhydron) | Sulfur: 350 mg (coal), carbon sulfur: 200 mg (coal) | Width: 46 cm x Depth: 79 cm x Height: 80 cm | 88 kg (with mount monitor) 78 kg (without mount monitor) |
Reviews shown here are reviews of companies.
Response Rate
100.0%
Response Time
126.8hrs