HACCP compatible chlorine concentration for drinks and food plant cleaning water management IR-31-27A/40/45/43ht/45ht monitorHandling Company
IWAKI CO., LTD.Product Image | Part Number | Price (excluding tax) | Contact output | Effective measurement range | Installation method | Measurement ingredient | Measurement method | Measurement principle | Repetition | Response time | Transmission output | Water inspection PH range | display | power consumption | power supply |
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
IR-31 |
Available upon quote |
Up, lower limit: 1A control output: 1A Hysteresis width can be selected by the control width setting slide switch |
5.0 ~ 50.0 mg/L or 30.0 ~ 300mg/L | Attach walls or 50A pipes (optional) | Effective chlorine | Bead cleaning / electrolytic cleaning combined type micro solid 3 electrode type | Electrochemical measurement method | Within ± 5%+1 digit of full scale (flow rate, pH, temperature, constant electric leader) | 90%Response from 1 minute to 1 minute from water inspection adjustment tank exit | DC4 ~ 20ma Insulated maximum load resistance 500Ω | PH5.8 ~ 10.5 (within a floating width of 1pH) |
Digital 3 digits LCD display Minimal resolution: 0.1 mg/L (FS 50.0 mg/L time) or 1mg/L (FS 300mg/L) |
Approximately 5VA | 1φac100 ~ 240V ± 10% 50/60Hz |