■ Temperature and humidity conformer using dry and wet bulb sensors ・ Temperature output: Outputs 0 to 20 MA DC or 4-20 MA DC for the dry cell input temperature 0-100 ° C or 0 to 200 ° C. ・ Humidity output: Outputs 0 to 20 MA DC or 4-20 MA DC for relative humidity 0 to 100 %RH. ■ Serial communication RS-485 is available optionally Compatible with our standard protocol, Modbus ASCII and Modbus RTU. You can configure communication parameters, read input / output, and read the state.

  • Product

    Temperature and humidity conformer THT-500




*Please note that we may not be able to accommodate sample requests.

1 Models of Temperature and humidity conformer THT-500

Product Image Part Number Price (excluding tax) Case material / color Conversion accuracy External dimensions Input dry ball input 1 Input dry ball input 2 Input sampling cycle Input wet bulb input Options (C5) Options (TC) Output -decomposition ability Output humidity output Output temperature output Power-supply voltage installation method mass power consumption usage environment
Temperature and humidity conformer THT-500-Part Number-THT-500

THT-500

Available upon quote Flame -retardant resin color: black ・ Temperature output: Temperature conversion within ± 1.0 ° C
・ Humidity output: Within ± 3 %of full scale
75 × 105 × 71 mm (W × H × D) Temperature resistance RTD PT100 3 conductor type 0-100 ° C (permissive input donductor resistance per line: 10Ω or less) Temperature resistance RTD PT100 3 conductor type 0 ~ 200 ° C (permissive input donductor resistance per line: 10Ω or less) (Dry ball can be switched to 0 to 200 ° C for 0 to 100 ° C in the instrument). 250ms Temperature resistance RTD PT100 3 conductor type 0-100 ° C (permissive input donductor resistance per line: 10Ω or less) Serial communication Terminal cover for electric shock prevention 12000 Output 0 to 20 ma DC or 4 to 20 mA DC (up to 550Ω) for 0 to 100%of the dry and wet bulb input temperature of 0 to 100 ° C for a dry ball input temperature of 0 to 100 ° C and wet bulb input temperature 0 to 100 ° C. Output 0 to 20mA DC or 4 to 20mA DC (maximum load resistance 550Ω) to 0 to 100 ° C or 0 to 200 ° C for dryball input temperature 100 ~ 240 V AC 50/60 Hz or 24 V AC/DC 50/60 Hz ・ The screws in the control panel (3 mm screws 4 places)
・ Installation angle: Steel plate (nickel plating)
Approximately 200g Approximately 5 VA ・ Viper temperature: 0 to 50 ℃
・ Viper humidity: 35-85 %RH (but not condensation)