Product
Measurement of trace gas components with particularly high reliability SS2100i Gas AnalyzerHandling Company
Endress+Hauser Group Services AGCategories
Product Image | Part Number | Price (excluding tax) | Analysis and measurement range | Analyzer adhesive material material | Channel | Dangerous place authentication | Housing material | Measured variable | Measurement principle | Movement pressure range | Peripheral temperature range | Product header | Product safety | Protection grade | communication | power supply |
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
SS2100i Gas Analyzer |
Available upon quote |
H2S (hydrogen sulfide): 0 ~ 10 ppmv 0 ~ 1000 ppmv 0 ~ 5000 ppmv 0 to 5% volume |
SUS 316L equivalent stainless steel FKM O ring Glass |
1 | IECEX / ATEX / CNEX / KC / CCOE / CML Zone 1 |
Electronic module: Bronze -free aluminum Laser/Celaine Close: Copper-free aluminum (I-2 only) Supply system enclosure: SUS 304 or 316 equivalent stainless steel |
concentration Cell pressure Cell temperature |
TDLAS |
Interior pressure: 140 ~ 350 kpa (20 ~ 50 PSIG) Samplecell: 800 ~ 1200 MBARA or 950 ~ 1700 MBARA (optional) |
-20 ~ 50 ° C (-4 ~ 122 ° f) Options (i-2 only): -10 ~ 60 ° C (14-140 ° F) |
Advanced gas analyzer for measurement of hydrogen sulfide (H2S). The SS2100i uses the TDLAS (tunable diode laser absorption spectroscope) to measure the concentration of the analytical object without physically contacting the fluid. | CE | IP66 |
Analog output: 2 x 4 ~ 20ma Insulation, maximum load 1200 Ω (DC 24 V) (measured value) Analog input: 1 x 4 ~ 20ma Insulation, maximum load 1200 Ω (DC 24 V) (process pressure) Serial: RS232C and Ethernet Protocol: Modbus Gould RTU, Daniel RTU or ASCII Digital output: 5, concentration alarm, general error, verification failure, verification 1 active, verification 2 active Digital input: 2, flow alarm, verification request |
AC 120 or 240 V ± 10%, 50 ~ 60 Hz; 300W |
Endress+Hauser Group Services AG