Water quality analysis parameters Digital dissolved oxygen sensor MEMOSENS COS22E-Memosens COS22E
Water quality analysis parameters Digital dissolved oxygen sensor MEMOSENS COS22E-Endress+Hauser Group Services AG

Water quality analysis parameters Digital dissolved oxygen sensor MEMOSENS COS22E
Endress+Hauser Group Services AG


About This Product

MEMOSENS 2.0 compatible aplvage oxygen sensor with Sanitary specifications ideal for food and life science industry ◾️ Application Memosens COS22E is perfect for Sanitari and sterilized applications for fully compliant with FDA and USP. Even in an environment where sterilization is frequent, it provides high long -term stability with minimal maintenance man -hours. The trace version protects your products, processes, and plant devices from the effects of dissolved oxygen. With Memosens 2.0 digital technology, COS22E provides platforms for preserving and efficient operation because Cos22E can save calibration and process data in more information. OXYMAX COS22E can measure the dissolved oxygen concentration in many industrial core processes. ・ Life Science and biotechnology: Process control in fermentation ・ Food and beverage industry ・ Water treatment: Boiler can water • Inactivate ・ Residual oxygen measurement in the process This digital sensor is available in dangerous places because it has obtained international explosion -proof authentication. ◾️Anucial ・ Reduction of maintenance man -hours Because the sensor has a module structure, you can quickly replace the diameter cap and electrolytic solution. ・ The safety of the product is guaranteed by complete compliance with all laws and regulations related to the sanitari process. ・ The highest level of product quality is achieved by accurate measurements that adopt high -precision measurement principles. ・ This sensor also supports dangerous places applications. ・ Compatible with industrial IoT Memosens 2.0 records proofreading and process data in more information, enables excellent trend identification, provides the basics of advanced industrial IoT services. ・ Improvement of plant occupancy rate Memosens technology and Liquiline converter are compatible with plug and play, so you can quickly set and replace sensors.

  • Product

    Water quality analysis parameters Digital dissolved oxygen sensor MEMOSENS COS22E




*Please note that we may not be able to accommodate sample requests.

1 Models of Water quality analysis parameters Digital dissolved oxygen sensor MEMOSENS COS22E

Product Image Part Number Price (excluding tax) Additional certification Characteristic External dimensions Material Measurement principle Measurement range Process pressure Process temperature Protection grade Temperature sensor application connection structure
Water quality analysis parameters Digital dissolved oxygen sensor MEMOSENS COS22E-Part Number-Memosens COS22E

Memosens COS22E

Available upon quote Additional certification Sanitor -type digital dissolved oxygen sensor, which lasts the maximum measurement stability in multiple sterile cycles, can be used in two measurement range Diameter: 12 mm (0.47 INCH)
Shaft length: 120, 160, 220, 320, 420 mm
(4.72, 6.30, 8.66, 12.60, 16.54 INCH)
Senshaft: Stainless steel 1.4435 (equivalent to SUS 316L)
Process Seal: FKM (USP <87>, <88> Class VI and FDA)
Explosion -proof process seal: FKM (FDA non -compliant)
Seal/O -ring: EPDM (USP <87>, <88> Class VI and FDA), FKM
(FDA)
Shaft sleeve: stainless steel 1.4435 (equivalent to SUS 316L) or titanium or hatery
The top layer of the septum: silicon (USP <87>, <88> class VI and FDA)
Squid film type dissolved oxygen measurement Standard sensor: 0 ~ 60 mg/L, 0 ~ 600 %SAT, 0 ~ 1200 HPA, 0 ~ 100 Vol %
A trace measurement sensor: 0 ~ 10 mg/L, 0 ~ 120 %SAT, 0 to 250 HPA, 0 to 25 voltes
0.1 ~ 1.2 mpa (14.5 ~ 174 PSI) Absolute pressure -5 ~ 135 ° C (23 ~ 275 ° f) IP68 NTC 22K Representative application:
-Pharmaceutical and biotechnology industry: process control and culture process control in enzyme production
.
-Water treatment: Boiler can water, WFI (injection water)
-Reggal oxygen measurement in the process
Non -contact connection head of electromagnetic induction transmission equipped with Memosens 2.0 technology structure

Customers who viewed this product also viewed

Other products of Endress+Hauser Group Services AG


View more products of Endress+Hauser Group Services AG

About Company Handling This Product

Endress+Hauser Group Services AG

  • United States of America

This is the version of our website addressed to speakers of English in the United States. If you are a resident of another country, please select the appropriate version of Metoree for your country in the drop-down menu.

Copyright © 2024 Metoree