Water quality analysis meter pH/ORP composite sensor MEMOSENS CPS76E corresponding to water quality analysis parameters-Memosens CPS76E
Water quality analysis meter pH/ORP composite sensor MEMOSENS CPS76E corresponding to water quality analysis parameters-Endress+Hauser Group Services AG

Water quality analysis meter pH/ORP composite sensor MEMOSENS CPS76E corresponding to water quality analysis parameters
Endress+Hauser Group Services AG


About This Product

MEMOSENS 2.0 compatible composite sensor that can be used for chemical processes and highly corrosive measurements ◾️ Application Memosens CPS76E is ideal for simultaneous measurement of PH/ORP in harsh environments. The unique contaminated reference guarantees a stable measurement not only for measurements with high contamination and corrosion, but also for measured materials with low conductivity. The Memosens 2.0 digital technology makes the CPS76E maximize the consistency of the process and the operation is simple. It has a moisture resistance, can be calibrated in the lab, and can save calibration and process data in more information, so we provide a foundation for preserving for predictive. Survival of the chemical industry and the following processes: ・ PH value or ORP value that changes rapidly ・ Liquid with high pollution to the electrode (eg, H2S) ATEX, IECEX, CSA C/US, NEPSI, domestic explosion, and Inmetro are acquired for use in the dangerous place Zone 0, 1, and 2. ◾️Anucial ・ Memosens 2.0 records proofreading and process data in more information, enables excellent trend identification, provides the basics of advanced industrial IoT services. ・ PH, ORP, RH value (in the case of RH mode) can be more detailed by simultaneous measurement of the process, so more strict process control is possible. -The optional pressurization reference provides highly reliable measurements for clogging, such as dispersion. ・ The unique ion wrap prevents the pollution of the liquid film of the electrode and the reference, and the long sensor life is guaranteed. -Pradent installation options can be installed with high flexibility. ・ The maximum process safety is ensured by non -contact electromagnetic induction signal transmission. ・ Minimize the downtime of the process and reduce operating costs by long sensor life.

  • Product

    Water quality analysis meter pH/ORP composite sensor MEMOSENS CPS76E corresponding to water quality analysis parameters




*Please note that we may not be able to accommodate sample requests.

1 Models of Water quality analysis meter pH/ORP composite sensor MEMOSENS CPS76E corresponding to water quality analysis parameters

Product Image Part Number Price (excluding tax) Characteristic Explosion -proof standard conformity certificate External dimensions Material Measurement principle Measurement range Process pressure Process temperature Protection grade Temperature sensor application connection structure
Water quality analysis meter pH/ORP composite sensor MEMOSENS CPS76E corresponding to water quality analysis parameters-Part Number-Memosens CPS76E

Memosens CPS76E

Available upon quote Digital PH/ORP electrode for chemical process applications with an ionized toxic reference Dangerous place ATEX, IECEX, CSA C/US, for use in Zone 0, Zone 1, Zone 2
NEPSI, domestic explosion, INMETRO certification
Diameter: 12 mm (0.46 INCH)
Shaft length: 120, 225, 360, 425 mm
(4.68, 8.77, 14.04, 16.57 INCH)
Senshaft: Glass suitable for processes
PH glass film: Type B, Type N
Metal lead: AG/AGCL
Ceramic liquid film, zirconium dinoxure
ORP measurement element: Shirokane
O -ring: FKM
Process coupling: PPS glass fiber strengthening
Nameplate: Metal Oxide Ceramic
Potential difference measurement Orp: --1,500 ~ 1,500 MV
Application B
・ PH: 0 ~ 14
Application H
・ PH: 0 ~ 12
Application B: 0.08 ~ 1.4 MPA (11.6 ~ 203 PSI) Absolute pressure
Application H: 0.08 ~ 0.7 MPA (11.6 ~ 101.5 PSI) Absolute pressure
Application B and H:
0 ~ 140 ° C (32-284 ° f)
Version TB:
0 ~ 140 ° C (32-284 ° f)
Version Tu, TP (Pressure Reference):
0 ~ 140 ° C (32-284 ° f) (140 ° C (284 ° F) only for sterilization)
In continuous operation, the pressure loss increases at a maximum of 100 ° C (212 ° f) and T> 100 ° C (212 ° f).
IP68 NTC 30k Process technology and the following process monitoring:
-PH value that changes rapidly
-Liquid with high pollution to the electrode (for example, H2S)
Digital connection head of electromagnetic induction transmission equipped with Memosens 2.0 technology With a temperature sensor with all shafts length
Advanced solid electrolyte (gel) technology

Customers who viewed this product also viewed

Other products of Endress+Hauser Group Services AG


View more products of Endress+Hauser Group Services AG

About Company Handling This Product

Endress+Hauser Group Services AG

  • United States of America

This is the version of our website addressed to speakers of English in the United States. If you are a resident of another country, please select the appropriate version of Metoree for your country in the drop-down menu.

Copyright © 2024 Metoree