Product
Multisizer 4E for measurement of particles and cells, volume, volume, and area granularity distributionHandling Company
Beckman Coulter Life SciencesProduct Image | Part Number | Price (excluding tax) | All count mode | Apartment diameter | Apature current intensity range | Current error of apartment | Dimensions / weight | Dynamic range (accuracy) | Electrolytic solution | Frequent count mode | Inject (quantitative) system | Installation conditions | Linearness | Measurement range | Number of pulse data | Particle size distribution data | Particle size distribution measurement range | Polar error | Power consumption | Processor | Quantification mode | Standard | Time setting mode | power supply | resolution |
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Multisizer 4E |
Available upon quote | 50 to 500,000 counts | 10 to 2,000 μm (nominal straight) | Can be set in 30 to 6,000 μA, 0.2 μA units | Within ± 0.4%of the set value | 640 (W) x 610 (D) x 510 (H) mm, unpacking weight: 45 kg (body) |
Particle diameter: 1: 40 (total), 1:30 (standard time) Particle volume: 1: 64,000 (total), 1:27,000 (standard time) |
Aquarium or organic solvent*can be used *Glass, fluororesin, fluorine elastomer, stainless steel solution |
10 to 100,000 counts | Injecting pump error to measure the volume by pouring a certain amount of suspension into the apartment: 0.5%or less |
Use this device in a closed space. Temperature: 5 to 40 ° C Humidity: 10 to 80%and no condensation |
Diameter: within ± 1% Volume: within ± 3% |
Extensions: 2-80% of the size of the apartment Standard range: 2 to 60% of the size of the apartment |
Up to 525,000 pulse in one measurement | Number of particle diameter, volume, absolute number of surface area, percent (%), particles concentration (/ml) |
Diameter: 0.2 ~ 1,600 μm Volume: 0.004 ~ 2.145 x 109 FL (μm3) |
0.5%or less | 55va Wat or less | High -speed signal digital processor | Can be set continuously in the range of 50 to 2,000 μL | Software: 21 CFR part 11 compliant |
Can be set in 0.1 to 999 seconds, 10 milliseconds. Usually set of 10 to 90 seconds |
AC single phase 100 ~ 120V ± 10%, 47 ~ 63 Hz, less than 210va (W), 3 or more ground needs | 4 to 400 channels (arbitrary settings possible), up to 11,000 channels (when ultra -highly decomposed) |