Sighting color meter SD7000-SD7000
Sighting color meter SD7000-Nippon Director

Sighting color meter SD7000
Nippon Director


About This Product

The SD 7000 type is a high -precision spectrosometer that can measure the wavelength range of 380 nm to 780nm in a 5nm interval output in reflection and transparent samples. The reflected sample is a measuring instrument that supports simultaneous measurements of (SCI) / excluding (SCE) and UV -cut filters, including (SCI) / (SCI) and (SCI). The display method can be selected according to the use of color management software using a PC connection, or from a dedicated label (CU-II) that uses a 5.7-inch large color LCD (CU-II) that is easy to use. Ideal for color management such as resin, glass, liquid crystal member, metallic, etc. ・ Measure the spectroscopy reflectance of the object and the spectroscopy transmittance. ・ Various samples such as solid, liquid, powder, pellet, and film can be measured. ・ The wavelength range from 380nm to 780 nm can be measured at 5 nm output wavelength intervals. ・ It can be measured (SCI) excluding (SCI) (SCI) (SCI). (Simultaneous measurement is also possible) ・ The power of the sample containing fluorescence (such as paper, fiber) is demonstrated by in / out in the UV -cut filter. ・ 5.7 -inch large color LCD is adopted on the display screen to display various graphs and approximation colors, making it easier to see. ・ Connect a USB memory to the measurement unit to save and read the measurement data (CSV file) and measurement conditions. -The measurement diameter changes according to the sample only change the lens by changing only the apartment. ・ The transparent sample room is made of a large sliding door system, making it easier to install samples and can measure large sizes. ・ Data on the PC screen and each graph can be analyzed by using color management software. (Optional)

  • Product

    Sighting color meter SD7000




*Please note that we may not be able to accommodate sample requests.

1 Models of Sighting color meter SD7000

Product Image Part Number Price (excluding tax) Lighting / light receiving conditions Measured diameter (lighting diameter) Measurement accuracy Measurement wavelength Power supply / power consumption Size / weight light source option
Sighting color meter SD7000-Part Number-SD7000

SD7000

Available upon quote Reflection/DI: 8 ° (diffusion lighting: 8 ° light receiving) Continuous reflected light (SCI) Reflection/DE: 8 ° (8 ° receiving light) Except for positive reflected light (SCE) transparent/DI: 0 ° (diffusion lighting: 0 ° light receiving) Reflection: LAVφ19mm (φ25.4mm), MAVφ8mm (φ12.7mm), SAVφ3mm (φ6.4mm) After stable time, a standard white plate 30 times at a standard white plate 30 times measurement spectroscopy reflectance: standard deviation 0.1 or less (⊿E*ab): Standard deviation 0.015 or more 380nm to 780nm5nm interval output AC100V -240V50/60Hz150VA W316mmXD462mmxh251mm about 14kg Pulse xenon lamp Various attachments

Customers who viewed this product also viewed

Other products of Nippon Director


View more products of Nippon Director

About Company Handling This Product

This is the version of our website addressed to speakers of English in the United States. If you are a resident of another country, please select the appropriate version of Metoree for your country in the drop-down menu.

Copyright © 2024 Metoree