This is a 250mm wide tabletop (flatbed) type recorder that supports 1, 2, and 3 channels.
It comes standard with a polarity switch, a chart paper fast forward switch, a chart rewind knob, a chart end alarm that activates when the chart paper runs out, and a superimposed event marker that allows you to place marks on the chart paper.
The standard specification is that the power supply is fixed at 100V AC, but if desired, it can be changed to a fixed voltage of 115V, 200V, 220V, or 240V.
A recording paper winder, recorder trolley, and separate pen-type event marker are available as options.
■Shape that makes it easy to write comments by hand
The length of the chart base plate has been lengthened to make it easier to write comments by hand over the recorded waveforms. In addition, there are no switches on the right sleeve panel so that they do not get in the way when handwriting comments.
■Built-in simple calibration power supply
It has a built-in power supply for calibration with the highest sensitivity of the input range.
Accuracy can be easily checked by following the steps below, without having to prepare a separate reference power source for calibration.
1. Set the "measurement range selector switch" on the input amplifier unit to the highest sensitivity.
2. Set the “zero point check switch” to “CHK” and set the zero point (0%).
3. When the "zero point check switch" is set to "CAL", the pen moves by the built-in calibration power supply, and if it stops at the full scale (100%) position, it is normal.
* Calibration is required if it deviates from the 100% position to + (plus) or – (minus).
■Chart paper feed operation section
Recording paper (chart paper) All settings related to recording paper (chart paper) can be made, such as switching the speed and switching the speed unit. The pen lift lever allows you to raise and lower the pen. Channel 1, channel 2, channel 3 can be operated individually. You can add a mark superimposed on the recorded waveform using the MARK switch.
■Input terminal
Three types of connections are possible: a signal line with the sheath removed, a signal line with a Y-type crimp terminal attached, and a signal line with a banana plug attached.