Process photometer (turbidity/SS concentration meter) NBP007-NBP007
Process photometer (turbidity/SS concentration meter) NBP007-Japan Laser Co., Ltd.

Process photometer (turbidity/SS concentration meter) NBP007
Japan Laser Co., Ltd.


About This Product

■Real-time measurement of process density and color ・High concentration SS concentration measurement 0.001% – 100% - Robust design with separate measuring head (cell) and measuring instrument ・Suspension liquids such as slurries and emulsions ・Measurement of density of cells and biomass ・Crystallization monitoring ・Control and optimization of CIP cleaning in place

  • Product

    Process photometer (turbidity/SS concentration meter) NBP007

Share this product


80+ people viewing

Last viewed: 14 hours ago


Free
Since our quotes are free, feel free to use our service.

No Phone Number Required
You won’t have to worry about receiving unnecessary calls.

1 Models of Process photometer (turbidity/SS concentration meter) NBP007

Product Image Part Number Price (excluding tax) Feature Software feature Features Automatic cleaning control Function Data Log Function Event log Function External input (digital) Function External output (analog) Function Power Function Relay output Function network Measurement performance Light source Measurement performance Measurement accuracy Measurement performance Measurement range Measurement performance Repeated measurement accuracy Measuring probe material Measuring probe piping connection Measuring probe pressure Measuring probe temperature Measuring probe window material
Process photometer (turbidity/SS concentration meter) NBP007-Part Number-NBP007

NBP007

Available upon quote Auto gain
Automatic zero point correction
Easy calibration function (save up to 16 products)
Manually or by external trigger
Automatic cleaning function
Over 17,000 points (timestamp, average, maximum & minimum)
ring buffer
Time interval: 1 second ~ 24 hours
16,000+ points, ring buffer
timestamp, alarm, zero point correction, cleaning,
Product changes, calibrations, system events
5 x digital inputs (voltage-free contacts)
Product/range setting, zero setting, cleaning, etc.
1 x 0 – 20 / 4 – 20 mA (NAMUR, max 21.6mA)
*Output can be added as an option
100-240VAC, 50-60Hz or 22-30VAC/VDC 1 x 1A 240VAC Failsafe
2 x 1A 240VAC User Defined (Alarm, PID, etc.)
1 x 1A 240VAC automatic cleaning control
Interface: Ethernet 10Base-T or 100Base-TX
Protocol: TCP/IP (HTML/Java, MODBUS)
Near infrared LED ± 2% (typical) Low: 0.0001 – 10%
High: 0.001 – 100%
*The range is set at the factory.
± 1% (typical) Stainless steel (SUS316L/EN1.4435)
or
Hastelloy C-22
Tri-Clamp (ISO 2852 & DIN 32676)
compatible DN25, DN32, DN40, 1″ and 1½”.
10mbar – 25bar (0.14 – 363psi) Up to 275℃ Sapphire