Plumbing type in-line turbidity meter TC007-TC007
Plumbing type in-line turbidity meter TC007-Japan Laser Co., Ltd.

Plumbing type in-line turbidity meter TC007
Japan Laser Co., Ltd.


About This Product

A turbidity meter that uses a light scattering principle and is equipped with a long-life LED. Robust design allows maintenance-free use even in harsh environments. ■Real-time & in-line turbidity measurement ・ISO 7027:1999 (E) compliant ratio method ・Measurement range: 0.01 to >4,000NTU/FTU - Robust design with separate measuring head (cell) and measuring instrument ・Maintenance free ・Many types of measurement cells that match many types of piping ・Analog output: 0/4-20 mA ・Modbus (slave) TCP/IP & Profibus (optional) ・Local & web-based GUI (TCP/IP) ・ATEX Exd IIB + H2 T6 IP66 Category II 2 G (optional) ■Simple & Robust Traditional turbidity meters are designed with sensitive parts such as light sources and photodetectors installed near the process line. As a result, measurement value drift and frequent calibration work are unavoidable under harsh process environments. On the other hand, Kemtrak's turbidity meter is mounted in a unit that separates sensitive parts from the measurement line. The signal light is transmitted via fiber, and a long-life LED is used as the light source. Kemtrak's products, which can be used without maintenance even in harsh environments, have received high praise all over the world.

  • Product

    Plumbing type in-line turbidity meter TC007

Share this product


110+ people viewing

Last viewed: 19 hours ago


Free
Since our quotes are free, feel free to use our service.

No Phone Number Required
You won’t have to worry about receiving unnecessary calls.

1 Models of Plumbing type in-line turbidity meter TC007

Product Image Part Number Price (excluding tax) Feature Software Features Automatic cleaning control Function Data Log Function Event log Function External input (digital) Function External output (analog) Function Power Function Relay output Function network Measurement performance Light source Measurement performance Measurement accuracy Measurement performance Measurement range Measurement performance Measurement resolution Measurement performance Repeated measurement accuracy
Plumbing type in-line turbidity meter TC007-Part Number-TC007

TC007

Available upon quote Auto gain
Automatic zero point correction
Easy calibration function (save up to 16 products)
Manually or by external trigger
Automatic cleaning function
Over 17,000 points (timestamp, average, maximum & minimum)
ring buffer
Time interval: 1 second ~ 24 hours
16,000+ points, ring buffer
timestamp, alarm, zero point correction, cleaning,
Product changes, calibrations, system events
5 x digital inputs (voltage-free contacts)
Product/range setting, zero setting, cleaning, etc.
1 x 0-20 / 4-20mA (NAMUR, max21.6mA)
*Output can be added as an option
100-240VAC, 50-60Hz or 22-30VAC/VDC 1 x 1A 240VAC Failsafe
2 x 1A 240VAC User Defined (Alarm, PID, etc.)
1 x 1A 240VAC automatic cleaning control
Interface: Ethernet 10Base-T or 100Base-TX
Protocol: TCP/IP (HTML/Java, MODBUS)
850nm
Other wavelengths are optional
±2% (typical) Ratio mode 0.01-1,000 NTU/FNU
Absorption mode 20-4,000 NTU/FNU
Backscatter Mode 100 – >5,000 NTU/FNU
0.01 at 0.01-10 NTU/FNU
0.1 at 10-100 NTU/FNU
1 at 100-1,000 NTU/FNU
10 at 1,000 – > 4,000 NTU/FNU
±1% or 0.01 NTU/FTU (typical)