Product
Strain calibrator CBM-122A/352AHandling Company
Tokyo Measuring Instruments Laboratory Co., Ltd.Categories
Items marked with have different values depending on the model number.
Click on the part number for more information about each product
Product Image | Part Number | Price (excluding tax) | Accuracy (when using the included input/output cable) AC bridge power supply | Accuracy (when using the included input/output cable) DC bridge power supply | Applicable bridge configuration | Applied voltage | Applied voltage Recommended | Applied voltage tolerance | Bridge resistance | Calibration value | Calibration value stability | External dimensions | Mass | Operating temperature and humidity range | Zero point stability |
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
CBM-122A |
Available upon quote | ± (0.02% of set value + 0.5×10^−6 strain) 23℃±3℃ | ± (0.02% of set value + 0.2×10^−6 strain) 23℃±3℃ | 4 gauge method | DC~20kHz | 3Vrms or less | 14Vp-p | 120Ω±1% | ±100~±10,900×10^−6 strain (100×10^−6 strain step), ±20,000, ±40,000×10^−6 strain | Less than ±9ppm/℃ of set value | 210 (W) × 84 (H) × 158 (D) mm | Approximately 1.4kg | 0~+50℃, 85%RH or less (excluding condensation) | ±0.8×10^−6 strain/℃ | |
CBM-352A |
Available upon quote | ± (0.02% of set value + 0.5×10^−6 strain) 23℃±3℃ | ± (0.02% of set value + 0.2×10^−6 strain) 23℃±3℃ | 4 gauge method | DC~20kHz | 5Vrms or less | 23Vp-p | 350Ω±1% | ±100~±10,900×10^−6 strain (100×10^−6 strain step), ±20,000, ±40,000×10^−6 strain | Less than ±6ppm/℃ of set value | 210 (W) × 84 (H) × 158 (D) mm | Approximately 1.4kg | 0~+50℃, 85%RH or less (excluding condensation) | ±0.8×10^−6 strain/℃ |
Click on the part number for more information about each product