Relatively Fast Response
Product
Strain calibrator CBA-131AHandling Company
Tokyo Measuring Instruments Laboratory Co., Ltd.Categories
Product Image | Part Number | Price (excluding tax) | AC power maximum power consumption | AC power supply allowable voltage | AC power supply rated voltage | Applicable bridge configuration | Bridge applied voltage | Bridge resistance | Calibration value 1000k range | Calibration value 100k range | Calibration value 10k range | Calibration value accuracy (23℃±3℃) Constant current DC bridge power supply 1 | Calibration value accuracy (23℃±3℃) Constant current DC bridge power supply 2 | Calibration value accuracy (23℃±3℃) Constant current DC bridge power supply 3 | Calibration value accuracy (23℃±3℃) Constant voltage 20kHz bridge power supply | Calibration value accuracy (23℃±3℃) Constant voltage 5kHz bridge power supply | Calibration value accuracy (23℃±3℃) Constant voltage DC bridge power supply 1 | Calibration value accuracy (23℃±3℃) Constant voltage DC bridge power supply 2 | Calibration value accuracy (23℃±3℃) Constant voltage DC bridge power supply 3 | Calibration value shift | Calibration value stability 20kHz bridge power supply | Calibration value stability 5kHz bridge power supply | Calibration value stability DC bridge power supply | Calibration voltage generator output resistance | Calibration voltage generator output voltage range | External dimensions | Mass | Operating temperature and humidity range | Output offset voltage (23℃±3℃) | Output offset voltage stability | Range switching |
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
CBA-131A |
Available upon quote | 15VA MAX | AC90~110V 50/60Hz | AC100V 50/60Hz | 4 gauge method |
・DC bridge power supply: 10V MAX ・5kHz bridge power supply: 5Vrms MAX ・20kHz bridge power supply: 2Vrms MAX |
・Input (A-C): 120Ω, 350Ω switchable ・Output (between B-D): 155Ω±65Ω |
± (100,010~1,000,000) ×10^−6 strain (Resolution 10×10^−6 strain) | ± (10,001~100,009) ×10^−6 strain (Resolution 1×10^−6 strain) | 0~±10,000.9×10^−6 strain (Resolution 0.1×10^−6 strain) |
・10k range (NIDS connector): ± (0.05% of setting+0.2×10^−6 strain) (Terminal block): ± (0.10% of setting+0.2×10^−6 strain) |
・100k range (NIDS connector): ± (0.05% of setting+2×10^−6 strain) (Terminal block): ± (0.10% of setting+2×10^−6 strain) |
・1,000k range (NIDS connector): ± (0.05% of setting+20×10^−6 strain) (Terminal block): ± (0.10% of setting+20×10^−6 strain) |
± (3% of setting+0.3×10^−6 strain) | ± (0.2% of setting+0.2×10^−6 strain) | 10k range: ± (0.01% of setting+0.2×10^−6 strain) | 100k: ± (0.01% of setting+2×10^−6 strain) | 1,000k: ± (0.01% of setting+20×10^−6 strain) | ±999.0×10^−6 strain | ± (200ppm of setting+0.3×10^−6 strain) /℃ | ± (20ppm of setting+0.05×10^−6 strain) /℃ | ± (10ppm of setting+0.02×10^−6 strain) /℃ |
・V-OUT: 0.5Ω MAX ・OUTPUT: 120Ω MAX |
0~±20V | 140 (W) × 100 (H) × 220 (D) mm | 2kg | 0~+50℃ |
・10k range: ±1μV ・100k range: ±2μV ・1,000k range: ±20μV |
・10k range: ±0.01μV/℃ ・100k range: ±0.1μV/℃ ・1,000k range: ±1μV/℃ |
Automatic |